Detail BATTERY ANALYZERS FLUKE BT500 SERIES
BATTERY ANALYZERS FLUKE BT500 SERIES
Spesification :
Functions Range Resolution Accuracy BT510 BT520 BT521
Battery resistance1 3 m© 0.001 m© 1 % + 8 Ï Ï Ï
30 m© 0.01 m© 0.8 % + 6 Ï Ï Ï
300 m© 0.1 m© 0.8 % + 6 Ï Ï Ï
3000 m© 1 m© 0.8 % + 6 Ï Ï Ï
Vdc 6 V 0.001 V 0.09 % + 5 Ï Ï Ï
60 V 0.01 V 0.09 % + 5 Ï Ï Ï
600 V 0.1 V 0.09 % + 5 Ï Ï Ï
1000 V 1 V 0.09 % + 5 Ï
Vac ( 45 Hz to 500 Hz with 800 Hz filter) 600 V 0.1 V 2 % + 10 Ï Ï Ï
Frequency ( displayed with Vac and Aac) 2 500 Hz 0.1 Hz 0.5 % + 8 Ï Ï Ï
AC voltage ripple ( 20 KHz Max) 600 mV 0.1 mV 3 % + 20 Ï Ï Ï
6000 mV 1 mV 3 % + 10 Ï Ï Ï
Adc/ Aac ( with accessory Fluke i410) 400 A 1 A 3.5 % + 2 Ï
Temperature 0 ° C to 60 ° C 1 ° C 2 ° C ( 4 ° F) Ï
Meter mode 999 records for each measurement position with time stamp
Sequence mode Up to 100 profiles and 100 profile templates ( Each profile stores up to 450 batteries) with time stamp
1The measurement is based on ac injection method. The injected source signal is < 100 mA, 1 kHz.
2 Trigger level VAC: 10 mV, Aac: 10 A
BT510 BT520 BT521
Resistance ( m© ) Ï Ï Ï
Battery voltage Ï Ï Ï
Voltage dc Ï Ï Ï
Voltage ac and frequency ( Hz) Ï Ï Ï
Ripple volt Ï Ï Ï
Temperature of negative battery post Ï
DC and ac current ( and frequency) Ï
DMM mode Ï Ï Ï
Sequence mode Ï Ï Ï
Discharge measurement mode Ï Ï Ï
Automatic measurement save Ï Ï Ï
Wireless communication Ï
Memory view Ï Ï Ï
Size ( HxWxD) 22 cm x 10.3 cm x 5.8 cm ( 9 in x 4 in x 2 in)
Weight 850 g ( 1.9 lb)
Screen dimensions 7.7 cm x 5.6 cm ( 3 in x 2.2 in)
Interface USB mini
Operating temperature 0 ° C to 40 ° C
Storage temperature -20 ° C to 50 ° C
Lithium-ion battery charging temperature 0 ° C to 40 ° C
Operating humidity Non-condensing ( 10 ° C)
< = 80 % RH ( at 10 ° C to 30 ° C)
< = 75 % RH ( at 30 ° C to 40 ° C)
Operating altitude Sea level to 2000 meters
Storage altitude Sea level to 12, 000 meters
IP rating IP40
Radio FCC Class A
Vibration requirements MIL-PRF-28800F: Class 2
Drop test requirements 1 meter
Temperature coefficients Add 0.1 x specified accuracy for each degree C above 28 ° C or below 18 ° C
Safety compliance 600 V CAT III
EMC IEC 61326
ROHS China, Europe
Protection Class 2 Pollution Degree II
Battery compliance UN38.3
UL2054
IEC62133
2G per IEC68-2-26, 25G, and 29
Tampilkan Lebih Banyak