Detail FLUKE 754 DOCUMENTING PROCESS CALIBRATOR
The Fluke 754 Series Documenting Process Calibrator with Integrated HART communication capability is the best Fluke process tools on the market. The Fluke 750 Series are the top of the line documenting process calibrators available to process professionals. The Fluke 754 combines all the process calibration tools into one powerful calibrator that does everything. Any process professional you speak with will know about Fluke Documenting Process Calibrators. They trust them to increase their troubleshooting and maintenance capabilities and will continue to do so with this improved model.
The Fluke 75-4 offers simultaneous source and measure capabilities for all common process parameters. Create and execute automated procedures and automatically capture the results. The USB interface enables two-way communication with popular PC-based instrumentation management applications.
The Fluke 754 goes even further by providing the ability to maintain and calibrate selected HART transmitters, a key process industry instrument. Any process professional you speak with will know about Fluke Documenting Process Calibrators. The Fluke 75/ 4 combines HART communication capability in a documenting process calibrator to deliver an integrated communicating calibrator. This rugged, reliable tool is ideal for calibrating, maintaining, and troubleshooting HART instrumentation.
Features:
Measure volts, mA, RTDs, thermocouples, frequency, and ohms to test sensors, transmitters and other instruments.
Source/ simulate volts, mA, thermocouples, RTDs, frequency, ohms, and pressure to calibrate transmitters.
Power transmitters during test using loop supply with simultaneous mA measurement.
Measure/ source pressure using any of 29 Fluke 700Pxx Pressure Modules.
Create and run automated as-found/ as-left procedures to satisfy quality programs or regulations. Record and document results.
Holds up to a full week of downloaded procedures and calibration results.
Use many features like autostep, custom units, user entered values during test, one-point and two-point switch testing, square root DP flow testing, programmable measurement delay etc.
Easy to use.
Three year warranty.
Bright white dual display. Read both sourced and measured parameters simultaneously.
Multi lingual interface.
Rechargeable NiMH battery for 10 hour uninterrupted use. Includes gas gauge.
Handling of fast pulsed RTD transmitters and PLCs, with pulses as short as 1 ms.
Delivered with DPC/ Track Sample software.
Compatible with many Asset Management software packages.
Includes:
Fluke 754 Documenting Process Calibrator ( HART)
Three sets of stackable test leads
Three sets of TP220
Two sets AC280 hook clips
BP7240 Li-ion battery pack
BC7240 battery charger
C799 field soft case
USB communication cable
Getting started guid
Instruction manual on CDROM
NIST traceable certificate of calibration
DPCTrack™ 2 sample software
HART communication cable.
Measurement Accuracy
Voltage DC
1 Year 2 Years
100.000 mV 0.02% + 0.005 mV 0.03% + 0.005 mV
3.00000 V 0.02% + 0.00005 V 0.03% + 0.00005 V
30.0000 V 0.02% + 0.0005 V 0.03% + 0.0005 V
300.00 V 0.05% + 0.05 V 0.07% + 0.05 V
Voltage AC
Range
40 to 500 Hz Resolution 1 Year 2 Years
3.000 V 0.001 V 0.5% + 0.002 V 1.0% + 0.004 V
30.00 V 0.01 V 0.5% + 0.02 V 1.0% + 0.04 V
300.0 V 0.1 V 0.5% + 0.2 V 1.0% + 0.2 V
Current DC
1 Year 2 Years
30.000 mA 0.01% + 5 uA 0.015% + 7 uA
110.00 mA 0.01% + 20 uA 0.015% + 30 uA
Resistance
1 Year 2 Years
10.000 ©
0.05% + 50 m© 0.07% + 70 m©
100.00 © 0.05% + 50 m© 0.07% + 70 m©
1.0000 k© 0.05% + 500 m© 0.07% + 0.5 ©
10.000 k© 0.1% + 10 © 0.15% + 15 ©
Frequency
Resolution Accuracy ( 2 Years)
1.00 to 110.00 Hz 0.01 Hz 0.05 Hz
110.1 to 1100.0 Hz 0.1 Hz 0.5 Hz
1.101 to 11.000 kHz 0.001 kHz 0.005 kHz
11.01 to 50.00 kHz 0.01 kHz 0.05 kHz
Source Accuracy
Voltage DC
1 Year 2 Years
100.000 mV
0.01% + 0.005 mV 0.015% + 0.005 mV
1.00000 V 0.01% + 0.00005 V 0.015% + 0.0005 V
15.0000 V 0.01% + 0.0005 V 0.015% + 0.0005 V
Current DC
1 Year 2 Years
22.000 mA ( Source) 0.01% + 0.003 mA 0.02% + 0.003 mA
Current Sink ( Simulate) 0.02% + 0.007 mA 0.04% + 0.007 mA
Resistance
1 Year 2 Years
10.000 ©
0.01% + 10 m© 0.015% + 15 m©
100.00 © 0.01% + 20 m© 0.015% + 30 m©
1.0000 k© 0.02% + 0.2 © 0.03% + 0.3 ©
10.000 k© 0.02% + 3 © 0.03% + 5 ©
Frequency
2 Years
0.1 to 10.99 Hz 0.01 Hz
0.01 to 10.99 Hz 0.01 Hz
11.00 to 109.99 Hz 0.1 Hz
110.0 to 1099.9 Hz 0.1 Hz
1.100 to 21.999 kHz 0.002 kHz
22.000 to 50.000 kHz
0.005 kHz
Tampilkan Lebih Banyak